NOVEL BIO-ELECTRODE FOR HUMAN BODY SIGNAL DETECTION 
G. S. Uthayakumar 
Associate Professor, St. Joseph’s Institute of Technology, Chennai, India 
ABSTRACT 


The current paradigm in healthcare is the notion of continuous remote patient monitoring using a network of wireless 
sensors. These healthcare sensor network systems, consisting of human body area networks (HBAN) and infrastructure 
area networks avoid the need for a manual self-administered health system and may enable users to take control of their 
health disorders in future. HBAN technology envisions miniaturized sensors worn or implanted on the body, continuously 
monitoring health parameters and acting to prevent the onset of critical health events. For example, diabetics now have 
access to an automatic insulin pump which monitors glucose levels and administers insulin when glucose levels are high. 
Similar technologies will one day result in devices which can minimize incidences of heart attack or stroke. They could 
prevent frequent hospital visits and save costs for both the individual patient and a nation’s healthcare system. Radio 
frequency (RF) wireless technology has been successfully deployed in most HBAN implementations; they consume a lot of 
battery power, susceptible to electromagnetic interference (EMI) and have security issues. The key issue with RF 
propagation in portable devices is that it consumes battery life quickly. For example zigbee has the maximum data rate of 
250 kb/s at 26.5mW resulting in 106nJ per received bit but IBC consumes an order of magnitude less energy at data rates 


up to IOMb/s which makes it an attractive communication method for HBAN applications. 


KEYWORDS: Wireless, Healthcare, Radio Frequency, Human Body, Networks, Security 


Article History 
Received: 31 May 2021 | Revised: 02 Jun 2021 | Accepted: 08 Jun 2021 


INTRODUCTION 


Intra body communication (IBC) is an alternative wireless communication technology which uses the human body as the 
signal propagation medium [1]. IBC has characteristics that could naturally address the issues with RF for BAN 
technology. This survey examines the on-going research in this area and highlights IBC core fundamentals [2], current 
mathematical models of the human body, IBC transceiver designs, and the remaining research challenges to be addressed. 
IBC has exciting prospects for making BAN technologies [4] more practical in the future. The IBC system is a protected 
and private communication network which provides natural security and interference free communication. The required 
operating frequency of IBC is much lower compared to RF system. This means signals are confined to the person’s 
proximity since reading data requires body contact. There is no signal leakage through the skin in IBC method and 
environmental noise has less effect on communication. At higher frequencies (300MHz to several gigahertz) the signal 
wavelength becomes comparable to the human body channel length and body radiates energy acting as an antenna. Since 
transmitter and receiver contain small size electrodes instead of antennas, the larger wavelength of the carrier signal 


compared to the electrode size results in interference free IBC below 300MHz. IBC forms a short range communication 


www.iaset.us editor @iaset.us 


8 G. S. Uthayakumar 


network inside and around human body and therefore allows the same frequency band to be reused by HBAN on other 
users with minimal interference. This property potentially allows designs to focus on improving data rates, reducing power 
consumption and integrating smaller form factors. IBC is classified into two basic coupling types called capacitive and 
galvanic coupling. For both coupling types the transceiver needs two pair of electrodes. In capacitive coupling only one of 
the electrodes of transmitter and receiver side is attached to the body, while other electrode is floating whereas in galvanic 
coupling both are attached to human body. The signal is generated between the body channel transceiver [6] by making a 
current loop through the external ground. The signal electrode of the transmitter induces the electric field into the human 
body which is controlled by the electric potential and body act as a conductor with the ground as the return path [7]. It is 
achieved by coupling AC current into human body. Here electrical signal is applied differentially between electrodes and a 
largely attenuated signal is received by the two receiver electrode [8]. In general, the ion content in the human body is the 


carrier of information in this method [9]. 
Electrical Properties of Human Body Tissues 


The two major properties are relative permittivity and electrical conductivity [10, 11]. Tissue types, the operational 
frequency range, temperature, intactness of cellular membranes and tissue water content are some of the major factors 
which determine the tissue electrical properties in the human body. Dielectric properties of living tissue vary differently 
with frequency dispersion which refers to the behavior of tissues at various frequency range; low frequency range[5], RF 
ranges and gigahertz frequency ranges, which are referred as alpha, beta and gamma distribution. Permittivity strongly 
declines, whereas conductivity increases within the frequency dispersion. The polarization of water molecules creates the 
gamma dispersion in the gigahertz region where the dispersion is not strong and has minimal effect on the electrical 
properties of body tissues. The polarization of cellular membranes is an obstacle for an ion to flow in or out of the cell and 
leads to the beta dispersion. The transport of ions across a biological membrane is related to the alpha dispersion where 
increase in tissue conductivity is rarely evident and permittivity shows a significant decrease. The existing systems used to 
improve security are biometric and password based systems. The wireless technology is used as a medium. This wireless 
data communication is based on radio frequency (RF) which have been successfully developed using popular protocols 
such as Bluetooth, Zigbee. A major drawback of wireless RF propagation for miniaturized monitoring devices is the high 
power consumption which limits the practical duration of an operation. Most current research claim that Zigbee and ANT 
has a battery life of three years, but this is at a low operating data rate, e.g., 1B transmitted per Smin. The IEEE 802.15.4 
standard for low power Zigbee protocol indicates a transmission power output of Odbm (1mW). Continuous operation at 
the maximum data rate of 250 kbps generally consumes a normal lithium ion battery in a matter of hours. It is evident that 
new approaches to ultralow power wireless technology are required for improving next generation BAN technology. The 
proposed system is used to improve the security using bio signals to access the door using a non-RF wireless data 
communication technique which uses human body as transmission medium for electrical signals which prevents energy 
being dissipated into the environment. It uses surface electrodes, microcontrollers and driver circuit. The data from the 
controller is converted to the electrical signal by using a driver circuit. The resultant signal is allowed to transmit through 
the body using surface electrodes. Another electrode is placed at different place of the body. It collects input from the 
human body and feeds the driver circuit. Driver and signal inducing circuit converts the electrical signal to the binarized 
input. The binarized input feeds the controller and it sends the command to the driver circuit. Depending upon the 


controllers command, devices will be switched on using driver circuit. 


Impact Factor (JCC): 7.1882 NAAS Rating 3.80 


Novel Bio-Electrode for Human Body Signal Detection 9 


EXPERIMENTAL SET UP 


The system consists of two nodes such as Transmitter and receiver node. The transmitter part consists of microcontroller, 
driver circuit, switches and electrode. The data are transmitted to the driver circuit by pressing the switch. Then the data’s 
are transmitted through the human body by surface electrodes. Switches which are interfaced with the microcontroller are 
used to start up the operation of transmission of data. The receiver part consists of microcontroller, driver circuit, relay and 
motor. This part is used to receive data from the human body by the surface electrodes. The data’s are received by the 
microcontroller using driver circuit. Depend upon the data from the microcontroller; driver circuit will give power supply 


to motor by using the relay 


Figure 1 & 2 shows the system consists of two nodes such as Transmitter and receiver node. The transmitter part 
consists of microcontroller, driver circuit, switches and electrode. The data are transmitted to the driver circuit by pressing 
the switch. Then the data’s are transmitted through the human body by surface electrodes. Switches which are interfaced 
with the microcontroller are used to start up the operation of transmission of data. The receiver part consists of 
microcontroller, driver circuit, relay and motor. This part is used to receive data from the human body by the surface 
electrodes. The data’s are received by the microcontroller using driver circuit. Depend upon the data from the 


microcontroller; driver circuit will give power supply to motor by using the relay. 


Driver and Surface 
MEMS Sensor - signalinducing |» — Bleotrode 
(Accelerometer. ——> Circuit 
sensor) ea 
| Switch 
| PICLES Human Body 


Power Supply 


Figure 1: Transmitter Block Diagram. 


Driver = | 
Surface rivet A|G | Rel 
Electrode [UNE [> pl | Pp Driver Ckt zy ee 
I 


Motor 
Human Body | pea 


Power Supply 


Figure 2: Receiver Block Diagram. 


www.iaset.us editor @iaset.us 


10 G. S. Uthayakumar 


Mems Accelerometer Adx1330 


The MEMS accelerometer contains a tiny cantilever beam with a proof / seismic mass. The sensor sends an analog signal 
as the proof mass deflects from its neutral position under the influence of external acceleration. MEMS accelerometer 
made by Analog Devices Inc., ADXL330 is used. This model already deploys multi-axes sensor in three-dimensional 
space. This provides a great advantage for the current interest since one does not need to set up at least two small MEMS 
accelerometers perpendicularly to obtain two dimensional orientations. Using ADXL330, it is able to measure the 
inclinations of object relative to two dimensional planes simultaneously. TAhe devise is also small in size, 4 x 4 x 1.45 
(mm) and consumes low power and integrates three-axes sensor in a single circuit. It also can measure both dynamic and 


static accelerations with enough accuracy that is within -3g and +3g, and sensitivity equal to 320 mV/ 1 gr. 
Asynchronous Receiver Transmitter (Usart) 


The Universal Synchronous Asynchronous Receiver Transmitter (USART) module is one of the two serial I/O modules. 
(USART is also known as a Serial Communications Interface or SCI). Bit SPEN and bits TRISC<7:6> have to be set in 
order to configure pins RC6/TX/CK and RC7/RX/DT as the Universal Synchronous Asynchronous Receiver Transmitter. 


USART Baud Rate Generator (BRG) 


The BRG supports both the Synchronous Asynchronous modes of the USART. It is a dedicated 8-bit baud rate generator. 
The SPBRG register controls the period of a free running 8-bit timer. In Synchronous Asynchronous mode, bit BRGH also 
controls the baud rate. Given the desired baud rate and FOSC, the nearest integer value for the SPBRG register can be 
calculated. From this, the error in baud rate can be determined. It may be advantageous to use the high baud rate (BRGH = 
1), even for slower baud clocks. This is because the FOSC/ (16(X + 1)) equation can reduce the baud rate error in some 
cases. Writing a new value to the SPBRG register causes the BRG timer to be reset (or cleared). This ensures the BRG 


does not wait for a timer overflow before outputting the new baud rate. 


The USART transmitter block diagram is shown in Figure 3.5. The heart of the transmitter is the transmit (serial) 
shift register (TSR). The shift register obtains its data from the read/write transmit buffer, TXREG. The TXREG register is 
loaded with data in software. The TSR register is not loaded until the STOP bit has been transmitted from the previous 
load. As soon as the STOP bit is transmitted, the TSR is loaded with new data from the TXREG register. Once the TXREG 
register transfers the data to the TSR register (occurs in one TCY), the TXREG register is empty and flag bit TXIF is set. 
This interrupt can be enabled / disabled by setting/clearing enable bit, TXIE 


f Data Bus 
TXIF }+ ———_ TXREG Register | 


TXIE 


Pin Buffer 
0 r_~| and Control 


i y 
Interrupt PERI OIEIRIO SSS RAS SASHES IS SAS ISIEAE SHS EIS 88 | RC6/TXICK pin 


TRMT | SPEN 


Figure 3: USART Baud Rate Generator. 


Impact Factor (JCC): 7.1882 NAAS Rating 3.80 


Novel Bio-Electrode for Human Body Signal Detection Il 


USART Receiver 


The data is received on the RC7/RX/DT pin and drives the data recovery block. Once Synchronous Asynchronous mode is 
selected, reception is enabled by setting bit CREN. The heart of the receiver is the receive (serial) shift register (RSR). 
After sampling the STOP bit, the received data in the RSR is transferred to the RCREG register (if it is empty). If the 


transfer is complete, flag bit RCIF is set. 


The actual interrupt can be enabled/ disabled by setting/clearing enable bit RCIE. Flag bit RCIF is a read only bit, 
which is cleared by the hardware. It is cleared when the RCREG register has been read and is empty. The RCREG is a 
double buffered register (i.e., it is a two deep FIFO). It is possible for two bytes of data to be received and transferred to the 
RCREG FIFO and a third byte to begin shifting to the RSR register. On the detection of the STOP bit of the third byte, if 
the RCREG register is still full, the overrun error bit OERR will be set. The word in the RSR will be lost. The RCREG 
register can be read twice to retrieve the two bytes in the FIFO. Overrun bit OERR has to be cleared in software. This is 


done by resetting the receive logic (CREN is cleared and then set). 


. « M84 Baud Rate CL ; [ o€RR | FERR | 
CREN 
—e SPBRG $| — ~~  — .LLcteccee | eas cee wae enwhe 
+64 ' MSb RSR Register LSb ' 
icpeansateecvuuet i ’ ' 
Baud Rate Generator L - STO | 8)| 7 | ece |! | 0 START ' 
' t ' 
RC7/RX/OT ‘ 


ees Cee eee eee eee eee er 


' 
f A 1 Pin Buffer - Data 
L \ and Contro’ Recovery 
l _ 


4 ' 
., RX9D) RCREG Register 
FIFO 
8 


wo terrunt {_} RCE Data Bus 
RCIE 


\ 


Figure 4: USART Receiver Block Diagram. 
Synchronous Asynchronous Reception 
When setting up a Synchronous Asynchronous Reception, follow these steps: 
e Initialize the SPBRG register for the appropriate baud rate. If a high speed baud rate is desired, set bit BRGH. 
USART Receiver 


The data is received on the RC7/RX/DT pin and drives the data recovery block. Once Synchronous Asynchronous mode is 
selected, reception is enabled by setting bit CREN. The heart of the receiver is the receive (serial) shift register (RSR). 
After sampling the STOP bit, the received data in the RSR is transferred to the RCREG register (if it is empty). If the 


transfer is complete, flag bit RCIF is set. 


The actual interrupt can be enabled/ disabled by setting/clearing enable bit RCIE. Flag bit RCIF is a read only bit, 
which is cleared by the hardware. It is cleared when the RCREG register has been read and is empty. The RCREG is a 
double buffered register (i.e., it is a two deep FIFO). It is possible for two bytes of data to be received and transferred to the 
RCREG FIFO and a third byte to begin shifting to the RSR register. On the detection of the STOP bit of the third byte, if 
the RCREG register is still full, the overrun error bit OERR will be set. The word in the RSR will be lost. The RCREG 
register can be read twice to retrieve the two bytes in the FIFO. Overrun bit OERR has to be cleared in software. This is 


done by resetting the receive logic (CREN is cleared and then set). 


www.iaset.us editor @iaset.us 


12 G. S. Uthayakumar 


Synchronous Asynchronous Reception 


When setting up a Synchronous Asynchronous Reception, follow these steps: Initialize the SPBRG register for the 


appropriate baud rate. If a high speed baud rate is desired, set bit BRGH. 
Surface Electrode 


The purpose of an electrode is to act as a connector between the subject’s skin (where electrical signals are easier to detect) 
and the MP40 acquisition unit (via the 40EL cable). If an electrode makes good contact with the skin, the signals that are 


generated will be relatively accurate. 


Electrodes are very simple devices that consist of a small piece of metal designed to make indirect contact with 
the skin and a larger adhesive plastic disk. Each electrode is about linch (2.5cm) in diameter and is sticky on one side so it 
will adhere to the skin. There is a small piece of plastic mesh filled with a bluish gel. Since gel conducts electricity (better 
than your skin) and is more flexible than the metal part of the electrode, the skin can flex and change shape somewhat 
without losing the electrical connection with the metal part of the electrode. BIOPAC disposable electrodes are standard 
electrodes which are widely used in clinical, research and teaching applications. These electrodes come in strips of ten and 


you should not remove an electrode from the backing until you are ready to use it. 
Electrode Placement 


There are two basic methods of electrode placement: monopolar and bipolar. In a monopolar recording, an active electrode 
is placed over the region of interest and a reference electrode is attached to a more distant part of the body. In a bipolar 
recording the voltage difference between two electrodes is measured with respect to the third “reference” electrode. Leads 


I, II, If are standard bipolar electrode configurations. The standard bipolar limb leads are 
Lead I = right arm (-), left arm (+) 
Lead II = right arm (-), left leg (+) 
Lead III = left arm (-), left leg (+) 

Preparing Electrode Site 


If signals are erratic, one way to improve electrode connections is to gently rub the area where the electrode is to be placed. 
This is known as abrading the skin and removes a thin layer of dead skin from the surface of the skin. Since dead skin 
doesn’t conduct electricity very well removing it improves the connection between the electrode and the skin. The skin site 


can be lightly abraded with a clean dry cloth. 
Attaching Electrodes 


To attach an electrode, peel the electrode from its backing and place it on the area. Once in place press down firmly on the 
electrode with two fingers and rock the electrode back and forth for a few seconds. This will ensure that it is adhering to 


the skin as much as possible. 


To help insure that the electrode will make good electrical contact with the skin, you may want to squeeze a 
drop or two of electrode gel onto either the surface of the skin or onto the electrode without allowing any to get on 


the adhesive. 


Impact Factor (JCC): 7.1882 NAAS Rating 3.80 


Novel Bio-Electrode for Human Body Signal Detection 13 


Connecting the Electrode Lead 


Each electrode lead cable is a different color and each pinch connector on the end of the cable needs to be attached to a 
specific electrode. The connector is polarized and needs to be clipped on such that the metal extensions inside the clip are 


on the down side to make surface contact with the electrode. 
Removing Electrodes 


Disconnect the electrode cable pinch connectors; peel the electrode off the skin and dispose of the electrodes. Wash the 
electrode gel residue from the skin, using soap and water. The electrodes may leave a slight ring on the skin for a few 


hours. This is normal and does not indicate that anything is wrong. 
RS232 Communication 


RS232 is an asynchronous serial communication protocol widely used in computers and digital systems. It is called 
asynchronous because there is no separate synchronizing clock signal as there are in other serial protocols like SPI and 
12C. The protocol is such that it automatically synchronizes itself. RS232 is easily used to create a data link between our 


MCU based projects and standard PC. 
Basics of Serial Communication 


In serial communication the whole data unit, say a byte is transmitted one bit at a time. While in parallel transmission the 
whole data unit, say a byte (8bits) are transmitted at once. Obviously serial transmission requires a single wire while 
parallel transfer requires as many wires as there are in our data unit. So parallel transfer is used to transfer data within short 
range while serial transfer is preferable in long range. In RS232 there are two data lines RX and TX. TX is the wire in 
which data is sent out to other device. RX is the line in which other device put the data it need to sent to the device. The 
arrow indicates the direction of data transfer. In addition to RX/TX lines there is a third line i.e. Ground (GND) or 
Common. As there is no “clock” line so for synchronization accurate timing is required so transmissions are carried out 
with certain standard speeds. The speeds are measured in bits per second. No of bits transmitted is also known as baud rate. 


Some standard baud rates are 1200, 2400, 4800, 9600. 
Rs232 Data Channels 


The data transfer is done in following ways, We make use of the following components PIC 16F877A Microcontroller, 
ULN2003, Relay, Surface electrodes, HD 74LSI4B. 


Figure 5 shows Using MEMS accelerometer the password (data) is sensed and with the axis value it enters the 
microcontroller. Switches interfaced with microcontroller start up the transmission operation only if the axis value is above 
the threshold range that is programmed in the controller. Crystal oscillator is used to provide clock frequency to 
microcontroller. Data are transmitted to the driver circuit by pressing the switch which converts binary to electrical signal, 
then transmitted through human body through surface electrodes. Since body has the electrical property it transmits it to the 
receiver side which receives data from the human body through the t surface electrodes which is placed at the other side of 
the body. Microcontroller receives data using driver circuit. Depending on the data, driver circuit will switch on/off the motor 
using relay. Here we make use of a device called ULN2003 between controller and relay because it needs 12V power supply 
to operate but the actual power obtained from the microcontroller will be of 5V. This device amplifies the power from 5V to 


12V. Thus from the output of the relay the motor switches on/off which indicates the opening or closing of the door. 


www.iaset.us editor @iaset.us 


14 G. S. Uthayakumar 


Receiver Section 
Transmitter Section “ s 


MEMS 
ACCELEROMETER 
SENSOR 


x 
z 
z 


Figure 5: Architecture Diagram. 


Implementation Using Software 


We make use of an embedded C language to program in the controller. The compiler used is CCS. 


Transmitter Simulation 


When the axis of the character is above 250, data will be transmitted. Transmission of the data is indicated by a LED (high). 


l= 
DA 


tep-nep 


Figure 6: Transmitter Simulation 1. 
When the axis of the character is below 250, data will not be transmitted. 


e ~=6©LED will not glow. 


e = It indicates the user to check the axis. 


Virtual Terminal 


5 
@ fhac 
ee 
Beeae 


¥ 


Figure 7: Transmitter Simulation 2. 


Impact Factor (JCC): 7.1882 NAAS Rating 3.80 


Novel Bio-Electrode for Human Body Signal Detection IS 


CONCLUSIONS 


This project has satisfactorily overcome the issues addressed by the RF wireless technology used in existing system. As we 


make use of human body as a medium the system becomes more efficient and advantageous. The security system is 


improved by making use of this concept since it provides natural security. As it consumes less power it increases the 


practical duration of the system. It provides an interference free communication and also there will be no signal leakage 


and it uses a lower frequency ranges. Due to all these reasons the security has been improved and its implementation is also 


simple. 
REFERENCES 
I. Giorgio Casinovi and Chad Young.” Estimation of power dissipation in switched-capacitor circuits. IEEE Trans. 


10. 


11, 


on CAD of Integrated Circuits and Systems, 22(12):1625—1636, 2003. 


D. Rakhmatov, S. Vrudhula, and D.A. Wallach. A model for battery lifetime analysis for organizing applications 
on a pocket computer. In Proc. IEEE Transactions on Very Large Scale Integration (VLSI) Systems, 2003 


D. Estrin, "Embedded Networked Sensing Research: Emerging System Challenges," in NSF Workshop on 


Distributed Communications and Signal Processing for Sensor Networks Evanston, Illinois, USA, 2002. 


D. Estrin, R. Govindan, J. Heidemann, and S. Kumar, "Next Century Challenges: Scalable Coordination in 
Sensor Networks," in IEEE/ACM International Conference on Mobile Computing and Networking, Seattle, 
Washington, USA, 1999, pp. 263-270. 


MEMS Technology page, URL: http://www.memsnet.org/ mems/what-is.html, Accessed in 10 April 2009. 


I. Khemapech, I. Duncan, and A. Miller, "A Survey of Wireless Sensor Networks Technology," in 6th Annual 
Postgraduate Symposium on the Convergence of Telecommunications, Networking and Broadcasting, Liverpool, 


UK, 2005. 


I. F. Akyldiz, W. Su, Y. Sankarasubramaniam, and E. Cayirci, "Wireless Sensor Networks: a Survey," Computer 
Networks (Elsevier), vol. 38, pp. 393-422, 2002. 


U. Varshney, "Pervasive Healthcare and Wireless Health Monitoring," Mobile Networks and Applications, vol. 
12, pp. 113-127, March 2007. 


K. W. Goh, J. Lavanya, Y. Kim, E. K. Tan, and C. B. Soh, "A PDA-based ECG Beat Detector for Home Cardiac 
Care," in IEEE Engineering in Medicine and Biology Society, Shanghai, China, 2005, pp. 375-378. 


H. Zhou, K. M. Hou, J. Ponsonnaille, L. Gineste, and C. D. Vaulx, "A Real-Time Continuous Cardiac 
Arrhythmias Detection System: RECAD," in IEEE Engineering in Medicine and Biology Society, Shanghai, 
China, 2005, pp. 875-881. 


H.-T. Chu, C.-C. Huang, Z.-H. Lian, and T. J. P. Tsai, "A Ubiquitous Warning System for Asthma Inducement," in 
IEEE International Conference on Sensor networks, Ubiquitous and Thrustworthy Computing, Taichung, Taiwan, 


2006, pp. 186-191. 